Evaluation of standardized porcine bone models to test primary stability of dental implants, using biomechanical tests and Micro-CT. An in vitro pilot study

  • C. Delmondes Freitas Dantas | info@ariesdue.it Department of Traumatology, Bucco-Maxillo-Facial Surgery and Periodontology, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil, Brazil.
  • D. Maeda Reino Department of Traumatology, Bucco-Maxillo-Facial Surgery and Periodontology, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil, Brazil.
  • J. Pires de Carvalho Department of Traumatology, Bucco-Maxillo-Facial Surgery and Periodontology, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil, Brazil.
  • A.B. Novaes Jr. Department of Traumatology, Bucco-Maxillo-Facial Surgery and Periodontology, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil, Brazil.
  • M. Taba Jr. Department of Traumatology, Bucco-Maxillo-Facial Surgery and Periodontology, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil, Brazil.
  • D. Bazan Palioto Department of Traumatology, Bucco-Maxillo-Facial Surgery and Periodontology, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil, Brazil.
  • A. Cândido dos Reis Department of Oral Rehabilitation, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil, Brazil.
  • S.L. Scombatti de Souza Department of Traumatology, Bucco-Maxillo-Facial Surgery and Periodontology, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil, Brazil.

Abstract

Aim This study evaluated a new porcine bone model to test the primary stability of different implants, analyzing Micro-CT, insertion torque, and pull-out strength.

Materials and methods Bone cylinders were prepared from porcine bone and separated into 2 groups: 10 high density bone cylinders (HDB), and 10 low density bone cylinders (LDB). Then, 3D pre-implant analyses were performed, evaluating tridimensional bone density (ratio of trabecular bone volume and total tomographic volume, BV/TV), trabecular separation; percentage of closed pores; percentage of open pores; percentage of total porosity, in 3 bone levels (L1 bone volume corresponding to the internal part of the threads; L2 corresponding to the area between 0 to 0.5 mm from the end of threads; L3 corresponding to the area between 0.5 to 1.5 mm from the end of threads). Twenty implants of two different macrostructures were inserted in the bone cylinders, and divided into 4 groups (5 implants each): Group 1, e-Fix HE implant placed in HDB cylinder; Group 2, e-Fix HE implant in LDB cylinder; Group 3, e-Fix HE Silver implant placed in HDB cylinder; Group 4, e-Fix HE Silver implant in LDB cylinder. The insertion torque was recorded and bone cylinders were re-evaluated by Micro-CT (post-implant analysis). Then a pull-out strength test was performed.

Results 3D analysis showed that pre- and post-implants intra-groups evaluation had statistically significant differences in Group 3 and 4, for all  tomographic parameters assessed. Group 3 showed the best values for biomechanical tests (Friedman Test, p<0.05).

Conclusion This methodology can produce standardized bone cylinders of high and low bone density, in which different implant designs are able to promote different effects, evidenced by biomechanical and image analysis.

Dimensions

Altmetric

PlumX Metrics

Downloads

Download data is not yet available.
Published
2015-10-30
Info
Issue
Section
Articles
Keywords:
Bone Density, Computerized tomography, Implant primary stability, Insertion torque, Pull-out strength
##article.stats##
  • Abstract views: 415

How to Cite
Delmondes Freitas Dantas, C., Maeda Reino, D., Pires de Carvalho, J., Novaes Jr., A., Taba Jr., M., Bazan Palioto, D., Cândido dos Reis, A., & Scombatti de Souza, S. (2015). Evaluation of standardized porcine bone models to test primary stability of dental implants, using biomechanical tests and Micro-CT. An in vitro pilot study. Journal of Osseointegration, 7(3), 43-51. https://doi.org/10.23805/jo.2015.07.03.01